FDA Adverse Event
Injury
Summary report: N
ICL (IMPLANTABLE COLLAMER LENS)
MDR report key: 1911723
·
Received November 17, 2010
Report
- Report Number
- 2023826-2010-01165
- Event Type
- Injury
- Date Received
- November 17, 2010
- Date of Event
- August 16, 2010
- Report Date
- October 22, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- MTA
- PMA / PMN Number
- P030016
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4) - SECONDARY SURGERY - (B)(4).
Description of Event or Problem · 1
THE REPORTER STATED THE SURGEON IMPLANTED A 12.5MM ICM125V4 IMPLANTABLE COLLAMER LENS ON (B)(6) 2010. THE LENS WAS EXPLANTED ON (B)(6) 2010 DUE TO EXCESSIVE VAULTING. SUBSEQUENTLY, THE PT EXPERIENCED NARROWING OF THE ANGLE AND SHALLOWING OF THE ANTERIOR CHAMBER. THE ICL WAS EXCHANGED FOR A SHORTER LENS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ICL (IMPLANTABLE COLLAMER LENS) | INTRAOCULAR LENS | MTA | STAAR SURGICAL CO. | ICM125V4 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR | Required Intervention | CARTRIDGE: MODEL AND LOT # UNK| FOAM TIP PLUNGER: MODEL AND LOT # UNK| INJECTOR: MODEL AND LOT # UNK |