FDA Adverse Event Malfunction Summary report: N

MAVERICK2 BALLOON CATHETER

MDR report key: 1911665 · Received December 1, 2010

Report

Report Number
2134265-2010-05265
Event Type
Malfunction
Date Received
December 1, 2010
Date of Event
November 5, 2010
Report Date
November 6, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE A FAILURE ANALYSIS OF THE DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED AND MODERATELY TORTUOUS MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE 15MM X 1.5MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED TO THE LESION FOR PRE-DILATATION AND THE BALLOON WAS INFLATED TO 6 ATMS. DURING THE SECOND INFLATION, THE BALLOON RUPTURED AT 10 ATMS. THE BALLOON WAS REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAVERICK2 BALLOON CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - GALWAY H7493892815150 12962908

Patients

Seq Age Sex Outcome Treatment
1