Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RECEIVED: RECEIVED = 1X X-SMART CARTRIDGE H1004C5210150. 1X X-SMART CASING ASSEMBLY H1004E2810010 SN: (B)(6). 1X X-SMART HEAD CAP H1004C5210500. 1X X-SMART NECK H1004C5470220. X-SMART CASING ASSEMBLY: SAV VARIOUS MECHANICAL PROBLEM; INSIDE OF THE HANDPIECE CASING IS BROKEN. X-SMART NECK: SAV BAD MAINTENANCE (USER); I NOTE THAT THE NECK IS RUSTED. X-SMART CARTRIDGE: SAV BAD MAINTENANCE (USER); I NOTE THAT THE CARTRIDGE IS RUSTED.

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION HAS CAUSED FILE SEPARATION. SINCE SEPARATION OF A FILE COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE WAS RETURNED AND INVESTIGATION RESULTS: AS PER SERVICE ENGINEER- H1004C5210500 - HEAD CAP -PROBLEM; H1004C5210150-CARTRIDGE- PROBLEM; H1004C5470220 - NECK- PROBLEM; H1004E2810010 -ASTING ASSEMBLY - PROBLEM; UNDER WARRANTY INVOICE NO. (B)(4),, DATED 12-12-22. THERE WAS NO INJURY TO THE PATIENT. PROBLEM INVESTIGATION-CARTRIDGE,NECK ,HEAD CAP AND CASTING ASSEMBLY -REPLACE CARTRIGDE ,NECK , HEAD CAP AND CASTING ASSEMBLY. 10.04.24: RMA COMMUNICATED TO DENTSPLY SIRONA INDIA + VIGILANCE.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A X-SMART W/CONTRA ANGLE EUR STOPPED DURING USE. NO INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1125849 X-SMART W/CONTRA ANGLE EUR HANDPIECE, DIRECT DRIVE, AC-POWERED EKX MAILLEFER INSTRUMENTS HOLDING SARL 0C250449

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown