X-SMART W/CONTRA ANGLE EUR
Report
- Report Number
- 8031010-2024-00039
- Event Type
- Malfunction
- Date Received
- April 16, 2024
- Date of Event
- April 10, 2024
- Report Date
- May 16, 2024
- Manufacturer
- MAILLEFER INSTRUMENTS HOLDING SARL
- Product Code
- EKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION RECEIVED: RECEIVED = 1X X-SMART CARTRIDGE H1004C5210150. 1X X-SMART CASING ASSEMBLY H1004E2810010 SN: (B)(6). 1X X-SMART HEAD CAP H1004C5210500. 1X X-SMART NECK H1004C5470220. X-SMART CASING ASSEMBLY: SAV VARIOUS MECHANICAL PROBLEM; INSIDE OF THE HANDPIECE CASING IS BROKEN. X-SMART NECK: SAV BAD MAINTENANCE (USER); I NOTE THAT THE NECK IS RUSTED. X-SMART CARTRIDGE: SAV BAD MAINTENANCE (USER); I NOTE THAT THE CARTRIDGE IS RUSTED.
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION HAS CAUSED FILE SEPARATION. SINCE SEPARATION OF A FILE COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE WAS RETURNED AND INVESTIGATION RESULTS: AS PER SERVICE ENGINEER- H1004C5210500 - HEAD CAP -PROBLEM; H1004C5210150-CARTRIDGE- PROBLEM; H1004C5470220 - NECK- PROBLEM; H1004E2810010 -ASTING ASSEMBLY - PROBLEM; UNDER WARRANTY INVOICE NO. (B)(4),, DATED 12-12-22. THERE WAS NO INJURY TO THE PATIENT. PROBLEM INVESTIGATION-CARTRIDGE,NECK ,HEAD CAP AND CASTING ASSEMBLY -REPLACE CARTRIGDE ,NECK , HEAD CAP AND CASTING ASSEMBLY. 10.04.24: RMA COMMUNICATED TO DENTSPLY SIRONA INDIA + VIGILANCE.
IN THIS EVENT IT IS REPORTED THAT A X-SMART W/CONTRA ANGLE EUR STOPPED DURING USE. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1125849 | X-SMART W/CONTRA ANGLE EUR | HANDPIECE, DIRECT DRIVE, AC-POWERED | EKX | MAILLEFER INSTRUMENTS HOLDING SARL | 0C250449 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |