CLEARCUT KNIVES
Report
- Report Number
- 2523835-2024-00650
- Event Type
- Malfunction
- Date Received
- April 16, 2024
- Date of Event
- March 13, 2024
- Report Date
- July 29, 2024
- Manufacturer
- ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
- Product Code
- HNN
- UDI-DI
- 00000000000000
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION IS PROVIDED IN SECTIONS H.6 AND H.11. A SAMPLE WAS NOT RECEIVED AT THE MANUFACTURING SITE FOR EVALUATION FOR THE REPORT OF BLUNT KNIVES; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. A REVIEW OF THE DEVICE HISTORY RECORD TRACEABLE TO THE REPORTED LOT NUMBERS, INDICATES THAT THE REPORTED PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE REPORTED PRODUCT¿S ACCEPTANCE CRITERIA. A SAMPLE WAS NOT RECEIVED AT THE MANUFACTURING SITE AND THE DEVICE HISTORY RECORD REVIEW OF THE LOT NUMBER PROVIDED INDICATED THE PRODUCT WAS PROCESSED AND RELEASED ACCORDING TO THE PRODUCT¿S ACCEPTABLE CRITERIA, THEREFORE; THE ROOT CAUSE FOR THE CUSTOMER COMPLAINT ISSUE CANNOT BE DETERMINED. A NONCONFORMANCE INVESTIGATION IS CURRENTLY IN PROGRESS TO INVESTIGATE THE TREND IDENTIFIED FOR DULL CUTTING INSTRUMENTS. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. COMPLAINTS ARE REVIEWED AND MONITORED AT REGULAR INTERVALS FOR ANY SIGNIFICANT ADVERSE TRENDS. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A NURSE REPORTED THAT DURING THE CATARACT SURGERY, OPHTHALMIC KNIVES WERE FOUND TO BE BLUNT. THE PROCEDURE WAS COMPLETED WITHOUT ANY HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 694930 | CLEARCUT KNIVES | KNIFE, OPHTHALMIC | HNN | ALCON RESEARCH, LLC - ALCON PRECISION DEVICE | NA | 167PW8 | 00000000000000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |