FDA Adverse Event Malfunction Summary report: N

APEX MONORAIL

MDR report key: 1911422 · Received December 1, 2010

Report

Report Number
2134265-2010-05124
Event Type
Malfunction
Date Received
December 1, 2010
Date of Event
November 4, 2010
Report Date
November 4, 2010
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL / PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE, A BALLOON RUPTURE OCCURRED. THE LESION WAS LOCATED IN A SEVERELY CALCIFIED CORONARY VESSEL. THE APEX MONORAIL 12MM X 2.00MM BALLOON WAS ADVANCED TO THE LESION AND INFLATED SEVERAL TIMES AND THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. THE PATIENT'S STATUS IS GOOD. FURTHER INFORMATION WAS REQUESTED BUT IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 APEX MONORAIL CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - GALWAY H7493895912200 13476834

Patients

Seq Age Sex Outcome Treatment