FDA Adverse Event Malfunction Summary report: N

MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 19113919 · Received April 16, 2024

Report

Report Number
2032227-2024-153090
Event Type
Malfunction
Date Received
April 16, 2024
Date of Event
March 23, 2024
Report Date
May 21, 2024
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000414344
PMA / PMN Number
P150001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

PUMP RECEIVED WITH BLANK DISPLAY. UNABLE TO VERIFY E/A ALARM UNSPECIFIED OR PERFORM THE DISPLACEMENT TEST, SLEEP CURRENT TEST, ACTIVE CURRENT TEST AND SELF TEST DUE TO BLANK DISPLAY. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, FORCE SENSOR, MOTOR, HARNESS ASSEMBLY VIBRATOR AND CORRODED BATTERY TUBE.2032227-2024-153090-1 TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED BATTERY TUBE, CRACKED BATTERY TUBE THREADS, CRACKED CASE AT CORNER OF THE BELT CLIP RAILS, CRACKED CASE ALONG THE SIDE OF THE BATTERY TUBE, PILLOWING KEYPAD OVERLAY AND MINOR SCRATCHED LCD WINDOW. BLANK DISPLAY DUE TO MOISTURE DAMAGE ON THE PCBA 1, PCBA 2, HARNESS ASSEMBLY AND CORRODED BATTERY TUBE. E/A ALARM UNSPECIFIED WAS UNKNOWN DUE TO BLANK DISPLAY. MOISTURE DAMAGE FOUND ON THE PCBA 1, PCBA 2, FORCE SENSOR, MOTOR, HARNESS ASSEMBLY VIBRATOR AND CORRODED BATTERY TUBE.¿ COSMETIC DAMAGE FOUND ON THE PUMP CASE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED A CRITICAL PUMP ERROR (OPEN BOOK IMAGE). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1880. TROUBLESHOOTING WAS PERFORMED AND THE CUSTOMER REPORTED AN OTHER CRITICAL PUMP ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE DEVICE. MMT-1880 WAS REQUESTED AND THE CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690723 MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1880 HG5GHCZZZ 000000763000414344

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male