MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP
Report
- Report Number
- 1423500-2010-06395
- Event Type
- Malfunction
- Date Received
- November 30, 2010
- Date of Event
- July 1, 2010
- Report Date
- July 19, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- PMA / PMN Number
- K882498
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT WAS CONFIRMED IN THE LAB FOR CRACK/BROKEN OCCLUDER FEET. THE SLEEVE OF THE USED SET WAS OPENED/CLOSED SEVERAL TIMES WITHOUT DIFFICULTY. FUNCTIONALLY HAND TIGHTENED A LAB TITANIUM ADAPTER WITHOUT DIFFICULTY. SET WAS THEN TESTED UNDERWATER AT 8 PSI WITH LEAK NOTED. DURING VISUAL INSPECTION, IT WAS NOTED THAT THERE WAS NO WHITISH SPOT ON THE OCCLUDER LEG, THE PLANT EVALUATION CITED NO VISIBLE SIGNS OF OVER-TORQUING. THE ROOT CAUSE OF THIS PROBLEM IS UNKNOWN. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT.
A CUSTOMER CONTACTED BAXTER TO REPORT THAT CRACKS WERE NOTED ON THE TWIST SWITCH OF THE TRANSFER SET AFTER (B)(6) OF USE. THERE WAS NO DISINFECTANT USED ON THE SET BY PATIENT OR DOCTOR. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED ALONG WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MINICAP EXTEND LIFE PD TRANSFSET W/TWIST CLAMP | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME | H09H25076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |