COLLEAGUE CXE VOLUMETRICINFUSION PUMP

MDR report key: 1911208 · Received November 30, 2010

Report

Report Number
6000001-2010-05345
Event Type
Malfunction
Date Received
November 30, 2010
Date of Event
June 22, 2010
Report Date
November 10, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION WAS CONFIRMED AND DUPLICATED. THE ASSIGNABLE CAUSE WAS DUE TO AN UIM (USER INTERFACE BOARD). THE UIM, MAIN BATTERIES AND HARNESS WERE REPLACED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). A SERVICE HISTORY REVIEW REVEALED THAT THE DEVICE HAS NOT BEEN PREVIOUSLY SERVICED BY BAXTER. A DEVICE HISTORY REVIEW HAS BEEN PERFORMED AND THERE WERE NO EXCEPTIONS NOTED DURING THE MANUFACTURE OF THIS LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED CONDITION. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING BAXTER'S REVIEW OF THE EVENT HISTORY FOR A SEPARATE ISSUE, IT WAS DISCOVERED THAT FAILURE CODE 511:320:658:0000 OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION ASSOCIATED WITH THIS REPORT IS 6.13.92, CATEGORIZED AS REMEDIATED.