FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL

MDR report key: 1911072 · Received November 30, 2010

Report

Report Number
3005075853-2010-06757
Event Type
Malfunction
Date Received
November 30, 2010
Date of Event
November 5, 2010
Report Date
November 8, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT 30 DEVICES WERE RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICES WERE FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. UPON EVALUATION OF THE DEVICES, THEY WERE FUNCTIONALLY LEAK TESTED AND PASSED. 29 DEVICES WERE FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. UPON EVALUATION OF ONE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. DUCKBILL, LEAK TEST FAILURE . BATCH # G9KH35 MFG DATE 7/07/2010; EXP DATE 6/07/2015; BATCH # G9KU20 MFG DATE 8/12/2010; EXP DATE 7/12/2015; BATCH # G9KE8Z MFG DATE 6/24/2010; EXP DATE 5/24/2015; BATCH # G9KH4J MFG DATE 7/08/2010; EXP DATE 6/08/2015; BATCH # G9JY96 MFG DATE 4/22/2010; EXP DATE 3/22/2015; BATCH # G9KK5M MFG DATE 7/16/2010; EXP DATE 2/22/2015.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AIR LEAKED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT. (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment
1