FDA Adverse Event Injury Summary report: N

GEMINI SELF-RETAINING CANNULA

MDR report key: 1910953 · Received November 23, 2010

Report

Report Number
1526605-2010-00002
Event Type
Injury
Date Received
November 23, 2010
Date of Event
October 21, 2010
Report Date
October 25, 2010
Manufacturer
FRANTZ MEDICAL DEVELOPMENT, LTD.
Product Code
NBH
PMA / PMN Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): VISUAL EVALUATION OF THE RETURNED DEVICE CONFIRMED THE WING AT THE DISTAL END OF THE CANNULA BROKE OFF DUE TO AN EXCESSIVE AXIAL LOAD. AXIAL LOAD TEST AND INVESTIGATION RESULTS FROM CANNULAS REPRESENTATIVE OF THE LOT CONFIRMED ACCEPTANCE TO THE PRODUCT SPECIFICATIONS. BASED ON THE INFORMATION PROVIDED AND EVALUATION OF THE RETURNED DEVICE, THE MOST LIKELY CAUSE OF THIS TYPE OF EVENT IS NOT DEPLOYING THE CANNULA ACCORDING TO THE INSTRUCTIONS. SPECIFICALLY, IT WAS IMPROPER PLACEMENT OF THE CANNULA IN THE SURGICAL SITE AND THE USE OF EXCESSIVE FORCE. GEMINI SELF-RETAINING CANNULA, INSTRUCTION FOR USE, WARNINGS AND PRECAUTIONS: APPLYING UNDO FORCE TO THE CANNULA ASSEMBLY AND/OR OBTURATOR DURING A PROCEDURE MAY DAMAGE THE PRODUCT AND/OR HARM THE PATIENT. WING DEPLOYMENT AND INNER CANNULA TELESCOPING SHOULD NOT REQUIRE EXCESSIVE FORCE.

Description of Event or Problem · 1

SURGEON PERFORMING ARTHROSCOPIC LABRAL REPAIR AND WAS USING GEMINI SR8 SELF RETAINING CANNULA SYSTEM. CANNULA HAS TWO SMALL WINGS AT THE END AND ONE WING BROKE OFF FOLLOWING REPAIR. XRAY WITH CARM UTILIZED BY SURGEON TO TRY AND LOCATE PLASTIC WING TIP. SURGEON SAW QUESTIONABLE SHADOW ON XRAY AND ATTEMPTED TO FIND WING TIP WITH OPEN EXPLORATION OF OPERATIVE SITE. ATTEMPT WAS UNSUCCESSFUL. PT STAYED OVERNIGHT IN HOSPITAL AND PROCEDURE WAS PROLONGED BY APPROXIMATELY 3.5 HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GEMINI SELF-RETAINING CANNULA NBH, ACCESSORIES, ARTHROSCOPIC NBH FRANTZ MEDICAL DEVELOPMENT, LTD. SR8 120901

Patients

Seq Age Sex Outcome Treatment
1 17 YR Hospitalization| R