OCTRODE PERCUTANEOUS LEAD
Report
- Report Number
- 1627487-2010-03342
- Event Type
- Injury
- Date Received
- November 23, 2010
- Date of Event
- October 27, 2010
- Report Date
- October 27, 2010
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION DIVISION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION: METHOD - THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS - THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. CONCLUSION - THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR. REPORT # 1627487-2010-03339. THE PATIENT WAS IMPLANTED WITH AN SCS SYSTEM ON (B)(6) 2010 CONSISTING OF AN IPG AND PERCUTANEOUS LEAD. IT WAS REPORTED THAT THE PATIENT IS EXPERIENCING OVERSTIMULATION AT THE IPG POCKET WHEN RECHARGING HIS IPG. THE REPORTED SENSATION IS ALSO FELT WHEN HE UTILIZED THE MAGNET TO TURN THE STIMULATION OFF. IN ADDITION, IT WAS REPORTED THAT HIS STIMULATION TURNS ON BY ITSELF. IN AN EFFORT TO RESOLVE THIS MATTER, A REPLACEMENT CHARGING SYSTEM WAS SHIPPED TO THE PATIENT. THE RESULTS OF THE INTERVENTION METHOD HAVE NOT BEEN DISCLOSED. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OCTRODE PERCUTANEOUS LEAD | SPINAL CORD STIMULATION LEAD | LGW | ST. JUDE MEDICAL - NEUROMODULATION DIVISION | 3186 | 3100540 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |