FDA Adverse Event
Malfunction
Summary report: N
CAREASSIST BED
MDR report key: 1910783
·
Received November 22, 2010
Report
- Report Number
- 9615739-2010-00273
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Date of Event
- October 29, 2010
- Report Date
- October 29, 2010
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND SIDECOMM CABLE CORROSION ON THE CONNECTION TO THE BED DUE TO FLUID AND A LOOSE PLATE ON THE CONNECTION END TO THE BIU. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE TO REPAIR THE BED.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THE BED WILL ONLY INTERMITTENTLY SEND NURSE CALL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAREASSIST BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |