R SERIES DEFIBRILLATOR
MDR report key: 1910754
·
Received November 19, 2010
Report
- Report Number
- 1220908-2010-03289
- Event Type
- Malfunction
- Date Received
- November 19, 2010
- Report Date
- October 28, 2010
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K060559
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORP EVALUATED THE DEVICE AND THE REPORTED MALFUNCTION WAS NOT REPLICATED OR CONFIRMED. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING, WHICH INCLUDED ENVIRONMENTAL AND FUNCTIONAL TESTING WITHOUT DUPLICATING THE MALFUNCTION. THE DIGITAL BOARD AND A SYS INTERCONNECT CABLE WERE REPLACED AS A PRECAUTION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS UNREADABLE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | R SERIES DEFIBRILLATOR | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | R SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |