FDA Adverse Event
Malfunction
Summary report: N
ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM
MDR report key: 19106100
·
Received April 15, 2024
Report
- Report Number
- 3006575795-2024-00108
- Event Type
- Malfunction
- Date Received
- April 15, 2024
- Date of Event
- March 22, 2024
- Report Date
- April 15, 2024
- Manufacturer
- ZYNO MEDICAL, LLC
- Product Code
- FRN
- UDI-DI
- 00814371020013
- PMA / PMN Number
- K130690
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
ON 3/22/2024 FLOWRATE ISSUE WAS OBSERVED DURING PREVENTIVE MAINTENANCE ACTIVITIES PERFORMED AT ZYNO MEDICAL, LLC. CAUSE NOT ESTABLISHED SPECIFICALLY. YEARLY PREVENTIVE MAINTENANCE ACTIVITIES PERFORMED ON THIS DEVICE.
Description of Event or Problem · 0
ON 3/22/2024 DURING SERVICING ACTIVITIES OF ZYNO MEDICAL, LLC WHICH INCLUDES PREVENTIVE MAINTENANCE THERE IS A FLOWRATE OFFSET% ISSUE WHERE FLOWRATE OFFSET% AT PRE-REWORK IS 9% AND AT POST-REWORK IT IS 2%. NO PATIENT INVOLVED AS THIS IS OBSERVED DURING PREVENTIVE MAINTENANCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 808438 | ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM | LARGE VOLUME INFUSION PUMP | FRN | ZYNO MEDICAL, LLC | Z800F | 20160608-SH | 00814371020013 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |