FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 1910515 · Received November 19, 2010

Report

Report Number
2953200-2010-02287
Event Type
Injury
Date Received
November 19, 2010
Date of Event
October 21, 2010
Report Date
October 21, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS AND CONCLUSION: ARTERIAL TRAUMA/DISSECTION/PERFORATION, SEVERELY TORTUOUS WITH 60 DEGREES BENDS IN THE VESSELS.

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM 1 MONTH AGO. (MFR 2953200-2010-02287) VESSEL MORPHOLOGY WAS REPORTED AS MILDLY CALCIFIED, AND SEVERELY TORTUOUS WITH 60 DEGREES BENDS. THERE WAS AN ILIAC ANEURYSM MEASURING 16 MM IN DIAMETER. IT WAS REPORTED THAT A RELIANT BALLOON WAS USED TO MODEL THE ILIAC STENT GRAFT. THE RELIANT BALLOON WAS WITHIN THE ILIAC STENT GRAFT WHEN IT WAS INFLATED WITH THE USE OF A 30 CC SYRINGE WITH 70% SALINE AND 30% CONTRAST SOLUTION AND UPON INFLATION THE ILIAC VESSEL WAS PERFORATED. (MFR 2953200-2010-02288) THE PHYSICIAN IMPLANTED AN ILIAC STENT GRAFT EXTENDING DOWN TO THE LEVEL OF THE HYPOGASTRIC ARTERY AND PERFORATION WAS RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00512078

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention