LAMITRODE TRIPOLE 16 SURGICAL LEAD

MDR report key: 1910343 · Received November 18, 2010

Report

Report Number
1627487-2010-03333
Event Type
Injury
Date Received
November 18, 2010
Date of Event
October 19, 2010
Report Date
October 19, 2010
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION DIVISION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL, METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DEVICE HISTORY RECORD FOUND A NONCONFORMANCE; HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

THE PT RECENTLY UNDERWENT AN IPG REPLACEMENT DUE TO OVERSTIMULATION (REF MFR REPORT # 1627487-2010-0327). HER ORIGINAL LEADS WERE CONNECTED TO THE NEW IPG; HOWEVER, IT WAS REPORTED THAT THE OVERSTIMULATION ISSUE PERSISTS. NO FURTHER INFO IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAMITRODE TRIPOLE 16 SURGICAL LEAD SPINAL CORD STIMULATION LEAD LGW ST JUDE MEDICAL - NEUROMODULATION DIVISION 3219 3093061

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention