FDA Adverse Event Injury Summary report: N

ENDO CLIP II MED/LRG 10MM PISTOL GRIP

MDR report key: 1910222 · Received November 22, 2010

Report

Report Number
1219930-2010-00885
Event Type
Injury
Date Received
November 22, 2010
Date of Event
October 18, 2010
Report Date
October 25, 2010
Manufacturer
UNITED STATES SURGICAL A
Product Code
FZP
PMA / PMN Number
K954435
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PROCEDURE: CHOLECYSTECTOMY. ACCORDING TO THE REPORTER: THE CLIPS DO NOT CLOSE PROPERLY AND SOME OF THE CLIPS CAME OUT TWISTED. THE SURGEON HAD TO TRANSECT THE CYSTIC DUCT AND ALSO HAD TO PLACE ADD'L CLIPPING IN ORDER TO CORRECT THE ISSUE. THERE WAS A SMALL BILE LEAK THAT WAS CONTROLLED WITHOUT ANY PROBLEMS. THE PROCEDURE WAS DELAYED MORE THAN THIRY MINUTES DUE TO THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDO CLIP II MED/LRG 10MM PISTOL GRIP DISPOSABLE CLIP APPLIER FZP UNITED STATES SURGICAL A N0F0663

Patients

Seq Age Sex Outcome Treatment
1 Other| S