FDA Adverse Event
Malfunction
Summary report: N
HARMONIC FOCUS CURVED SHEARS, TORQUE WRENCH AND GRIP ASSIST
MDR report key: 1910117
·
Received February 5, 2008
Report
- Report Number
- 3005075853-2008-00014
- Event Type
- Malfunction
- Date Received
- February 5, 2008
- Date of Event
- January 10, 2008
- Report Date
- January 15, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K063192
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). INFO ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED DURING A THYROID PROCEDURE, THE DEVICE WAS CHIRPING MORE THAN NORMAL AND CUTTING VERY SLOW. THEY USED ANOTHER LIKE DEVICE TO COMPLETE CASE WITH NO PT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC FOCUS CURVED SHEARS, TORQUE WRENCH AND GRIP ASSIST | LFL | ETHICON ENDO-SURGERY, LLC | NA | D4JK2X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |