FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1910029 · Received November 10, 2010

Report

Report Number
1720753-2010-04027
Event Type
Malfunction
Date Received
November 10, 2010
Date of Event
October 21, 2010
Report Date
November 10, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE TEMPERATURE SENSOR WAS REPLACED AND THE HIGH VOLTAGE CABLE WAS CLEANED AND DEGREASED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 9800 SYSTEM WAS ARCING AND HAD A TEMPERATURE SENSOR ERROR DURING A CASE. THE PROCEDURE WAS COMPLETED. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1