FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1910023 · Received November 10, 2010

Report

Report Number
1720753-2010-04032
Event Type
Malfunction
Date Received
November 10, 2010
Date of Event
October 22, 2010
Report Date
November 10, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND PERFORMED A CALIBRATION OF THE COLLIMATOR STOP. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYSTEM DISPLAYS AN IRIS POTENTIOMETER ERROR UPON BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1