FDA Adverse Event Malfunction Summary report: N

CARDINAL HEALTH

MDR report key: 19096580 · Received April 12, 2024

Report

Report Number
3005670221-2024-00023
Event Type
Malfunction
Date Received
April 12, 2024
Report Date
April 12, 2024
Manufacturer
JIANGSU CAINA MEDICAL CO., LTD.
Product Code
PNR
UDI-DI
06971564466202
PMA / PMN Number
K190502
Removal / Correction Number
Z-0858-2024
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

RETURNED SAMPLE WAS REQUESTED BUT HAVEN'T RECEIVED YET. THE DHR OF THIS LOT WAS REVIEWED WITHOUT ANY ISSUE. THE DESIGN HISTORY FILE AND 510K SUMBSSION WERE REVIEWED. THIS SYRINGE IS DESIGNED FOR MANUAL USE ONLY, NOT FOR PUMP USE. FOR MANUAL USE OR NOT FOR PUMP USE WAS NOT PRESENTED ON ANY LABELING OF THIS PRODUCT. THE COMPLIANT HISTORY WAS REVIEWED THREE PUMP COMPATBILITY ISSUES WITH THIS PRODUCT WERE RECEIVED BEFORE. A RECALL WAS CONDUCTED BY DISTRIBUTOR CARDINAL HEALTH 200 LLC WITH RECALL NUMBER Z-0858-2024. THE CORRETIVE ACTION PUTTING THE CAUTION "FOR MANUAL USE OR NOT FOR PUMP USE" ON ALL SYRINGES' LABELING OF K190503 WILL BE TAKEN. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ANY ADDITIONAL INFORMATION FROM THE RECEIVED SAMPLES.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE SYRINGES CANNOT BE USED WITH THE FEEDING PUMPS. THERE WAS NO HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
793904 CARDINAL HEALTH ENFIF ENTERAL SYRINGE PNR JIANGSU CAINA MEDICAL CO., LTD. 230701 06971564466202

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other