FDA Adverse Event Malfunction Summary report: N

NIMBUS H AMBULATORY INFUSION PUMP

MDR report key: 19095260 · Received April 12, 2024

Report

Report Number
3011581906-2024-00530
Event Type
Malfunction
Date Received
April 12, 2024
Date of Event
October 15, 2021
Report Date
April 12, 2024
Manufacturer
INFUTRONIX, LLC
Product Code
FRN
UDI-DI
00817170020086
PMA / PMN Number
K153193
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INFUTRONIX ISSUED A PRODUCT QUALITY HOLD ORDER TO VERIFY ALL THE PRODUCTS IN THE INVENTORY ARE FREE FROM SIMILAR ISSUES THREE CUSTOMER LETTERS WERE SENT TO AFFECTED CUSTOMERS (B)(6) HOSPITAL. UNRELEASED VERSION 4 DUKE PROTOCOL LIBRARY INSTALLED ON THE NIMBUS PAINPRO (H) PUMP BY THE CUSTOMERS' SERVICE PROVIDER INSTEAD OF THE INTENDED DUKE V3 PROTOCOL LIBRARY INFUTRONIX ISSUED A HEALTH HAZARD EVALUATION INFUTRONIX ISSUED A FIELD ACTION PLAN TO ADDRESS THIS ISSUE. THIS FIELD ACTION WILL REPLACE THE AFFECTED INFUSION PUMP WITH PUMPS INSTALLED WITH THE CORRECT DUKE V3 PROTOCOL LIBRARY. DHR WAS REVIEWED, AND THE PUMP PASSED ALL PREVIOUS TESTS. RETURNED COMPLAINT DEVICES ARE NOT AVAILABLE AS OF 4/12/2024 OR RECERTIFIED POST-INVESTIGATION. INTENDED LIBRARY: DUKE V3. ACTUAL LIBRARY: DUKE V4 (UNRELEASED) BOTH VERSIONS OF THE DUKE LIBRARY SHARE THE SAME CLINICIAN CODE "191". THE DEVICE WAS OPERABLE BY THE USER.

Description of Event or Problem · 0

ON 10/19/2021 A COMPLAINT WAS OPENED IN QOS INTUVIE RECEIVED A COMPLAINT THAT THE PUMP HAS INCORRECT LIBRARY CONFIGURATION. UNRELEASED VERSION 4 DUKE PROTOCOL LIBRARY INSTALLED ON THE NIMBUS PAINPRO (H) PUMP BY THE CUSTOMERS' SERVICE PROVIDER INSTEAD OF THE INTENDED DUKE V3 PROTOCOL LIBRARY. NO OTHER DETAILS WERE PROVIDED ASSOCIATED WITH THE EVENT. NO PATIENT HARM WAS REPORTED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1413001 NIMBUS H AMBULATORY INFUSION PUMP AMBULATORY INFUSION PUMP FRN INFUTRONIX, LLC NIMBUS H 2009231440 00817170020086

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown