FDA Adverse Event Malfunction Summary report: N

NIMBUS II AMBULATORY INFUSION PUMP

MDR report key: 19095159 · Received April 12, 2024

Report

Report Number
3011581906-2024-00306
Event Type
Malfunction
Date Received
April 12, 2024
Date of Event
September 21, 2020
Report Date
April 12, 2024
Manufacturer
INFUTRONIX, LLC
Product Code
FRN
PMA / PMN Number
K153193
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF DEVICE HISTORY RECORD AND REVIEW OF PRODUCTION RECORDS CANNOT BE PERFORMED AS THERE IS NO SERIAL NUMBER OF THE PUMP MENTIONED WITHIN THE COMPLAINT RECEIVED. ALSO COMPLAINT DATA CANNOT BE REVIEWED ON THIS DEVICE. THIS IS A COMPLAINT FROM THE YEAR 2020. THE FOLLOW UP MDR WILL BE CREATED WHEN WE HAVE ADDITIONAL INFORMATION AVAILABLE ON THE PUMP. INVESTIGATION CANNOT BE PERFORMED AS PUMP IS NOT AVAILABLE FOR EVALUATION. A CAPA HAS BEEN OPENED IN ORDER TO FULLY DIAGNOSE AND ADDRESS THE ROOT CAUSE OF THE REPORTED EVENT. INTENDED LIBRARY: "STANFORD". ACTUAL LIBRARY: STFD 2 V3. THE SPECIFIC INTENDED LIBRARY IS NOT CLEAR BASED ON THE COMPLAINT FILE, ONLY THAT IT WAS INTENDED FOR STANFORD AND THE LIBRARY ON THE DEVICE RECEIVED WAS REPORTED AS INCORRECT. THE STFD 2 V3 LIBRARY DOES NOT HAVE A CLINICIAN CODE. THE DEVICE WAS OPERABLE BY THE USER.

Description of Event or Problem · 0

ON 09/21/2020 A COMPLAINT WAS OPENED IN QOS INTUVIE RECEIVED A COMPLAINT THAT THE PUMP HAS INCORRECT LIBRARY CONFIGURATION. NO OTHER DETAILS WERE PROVIDED ASSOCIATED WITH THE EVENT. NO PATIENT HARM WAS REPORTED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1401562 NIMBUS II AMBULATORY INFUSION PUMP AMBULATORY INFUSION PUMP FRN INFUTRONIX, LLC N/I N/I

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown