NIMBUS II AMBULATORY INFUSION PUMP
Report
- Report Number
- 3011581906-2024-00298
- Event Type
- Malfunction
- Date Received
- April 12, 2024
- Date of Event
- September 21, 2020
- Report Date
- April 12, 2024
- Manufacturer
- INFUTRONIX, LLC
- Product Code
- FRN
- PMA / PMN Number
- K153193
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
A REVIEW OF DEVICE HISTORY RECORD AND REVIEW OF PRODUCTION RECORDS CANNOT BE PERFORMED AS THERE IS NO SERIAL NUMBER OF THE PUMP MENTIONED WITHIN THE COMPLAINT RECEIVED. ALSO COMPLAINT DATA CANNOT BE REVIEWED ON THIS DEVICE. THIS IS A COMPLAINT FROM THE YEAR 2020. THE FOLLOW UP MDR WILL BE CREATED WHEN WE HAVE ADDITIONAL INFORMATION AVAILABLE ON THE PUMP. INVESTIGATION CANNOT BE PERFORMED AS PUMP IS NOT AVAILABLE FOR EVALUATION. A CAPA HAS BEEN OPENED IN ORDER TO FULLY DIAGNOSE AND ADDRESS THE ROOT CAUSE OF THE REPORTED EVENT. INTENDED LIBRARY: HMC V2 (135). ACTUAL LIBRARY: DEFAULT . THE DEFAULT LIBRARY DOES NOT HAVE A CLINICIAN CODE. THE DEVICE WAS OPERABLE BY THE USER.
ON 09/21/2020 A COMPLAINT WAS OPENED IN QOS INTUVIE RECEIVED A COMPLAINT THAT THE PUMP HAS INCORRECT LIBRARY CONFIGURATION. NO OTHER DETAILS WERE PROVIDED ASSOCIATED WITH THE EVENT. NO PATIENT HARM WAS REPORTED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 837039 | NIMBUS II AMBULATORY INFUSION PUMP | AMBULATORY INFUSION PUMP | FRN | INFUTRONIX, LLC | N/I | N/I |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |