FDA Adverse Event Malfunction Summary report: N

CARDINALHEALTH

MDR report key: 19094159 · Received April 11, 2024

Report

Report Number
3005670221-2024-00008
Event Type
Malfunction
Date Received
April 11, 2024
Report Date
April 12, 2024
Manufacturer
JIANGSU CAINA MEDICAL CO., LTD.
Product Code
FMF
UDI-DI
10192253034530
PMA / PMN Number
K113091
Removal / Correction Number
Z-0851-2024
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

RETURNED SAMPLE WAS REQUESTED BUT HAVEN'T RECEIVED YET. THE DHR OF THIS LOT WAS REVIEWED WITHOUT ANY ISSUE. THE DESIGN HISTORY FILE AND 510K SUMBSSION WERE REVIEWED. THIS SYRINGE IS DESIGNED FOR MANUAL USE ONLY, NOT FOR PUMP USE. FOR MANUAL USE OR NOT FOR PUMP USE WAS NOT PRESENTED ON ANY LABELING OF THIS PRODUCT. THE COMPLIANT HISTORY WAS REVIEWED NO PUMP COMPATBILITY ISSUES WITH THIS PRODUCT WERE RECEIVED BEFORE. A RECALL WAS CONDUCTED BY DISTRIBUTOR CARDINAL HEALTH 200 LLC WITH RECALL NUMBER Z-0851-2024. THE CORRETIVE ACTION PUTTING THE CAUTION "FOR MANUAL USE OR NOT FOR PUMP USE" ON ALL SYRINGES' LABELING OF K113091 WILL BE TAKEN. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ANY ADDITIONAL INFORMATION FROM THE RECEIVED SAMPLES.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THEY RECEIVED SOME FUNNY-LOOKING TB SYRINGES WITH AN EXTENDED BLACK PLUNGER TIP THAT LOOKS LIKE THE TIP OF A SHARPIE. THE INCORRECT ITEM CAUSED SOME ISSUES IN THEIR CARDIAC AREA WITH THEM NOT BEING ABLE TO FULLY INFUSE MEDICATIONS. PHOTOS ATTACHED. ADDITIONAL INFORMATION RECEIVED ON 09AUG2023 STATED THAT THERE WAS NO INJURY TO THE PATIENT BUT DID RESULT IN A DELAY IN MEDICATION ADMINISTRATION. NO INTERVENTIONS WERE REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
836974 CARDINALHEALTH SYRINGE FMF JIANGSU CAINA MEDICAL CO., LTD. 230201 230204 10192253034530

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other