FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III XENON LIGHT SOURCE

MDR report key: 19094155 · Received April 11, 2024

Report

Report Number
3002808148-2024-03428
Event Type
Malfunction
Date Received
April 11, 2024
Date of Event
March 26, 2024
Report Date
December 18, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
NWB
UDI-DI
04953170298868
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, OLYMPUS CONFIRMED THE REPORTED EVENT. THE LOCK MECHANISM AND THE SCOPE DETECTION SLIDE ARE NOT FUNCTIONING, AND FROM THIS, WE PRESUME THAT THE FRONT PANEL BLINKED. HOWEVER, THE PRODUCT HAS NOT BEEN RETURNED TO QAD OF SHIRAKAWA PLANT AND WE CANNOT CHECK THE PRODUCT¿S DETAILED STATUS, THUS WE COULD NOT DETERMINE THE ROOT CAUSE. ABOUT ASSEMBLING NON-OLYMPUS PRODUCT, WE FOUND THE DESCRIPTION BELOW IN THE INSTRUCTION MANUAL. WARNING NEVER INSTALL A LAMP THAT HAS NOT BEEN APPROVED BY OLYMPUS. THE USE OF A NONAPPROVED LAMP CAN CAUSE DAMAGE TO THE LIGHT SOURCE AND ANCILLARY EQUIPMENT, MALFUNCTION OR A FIRE. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THE XENON LIGHT SOURCE HAD CORROSION ON SOCKET PINS AND HAD FRONT PANEL MOUTH DAMAGED. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
836970 EVIS EXERA III XENON LIGHT SOURCE XENON LIGHT SOURCE NWB SHIRAKAWA OLYMPUS CO., LTD. CLV-190 04953170298868

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown