PALODENT PLUS PIN TWEEZERS
Report
- Report Number
- 2515379-2024-00018
- Event Type
- Malfunction
- Date Received
- April 11, 2024
- Report Date
- May 10, 2024
- Manufacturer
- DENTSPLY LLC
- Product Code
- DZN
- UDI-DI
- D0026598201
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
INVESTIGATION RETURN 5-9-2024: RETURNED PRODUCT 1 PAIR OF PALODENT PLUS PIN TWEEZERS ITEM# 759820 BATCH# 100322 WAS INSPECTED AND CONFIRMED TO BE MISSING THE ¿PIN¿ AT THE TIPS OF THE TWEEZERS WHICH AIDS IN HOLDING THE TAB MATRICES IN PLACE WITHIN THE INSTRUMENT. THE RETURNED PRODUCT IS NOT TO SPECIFICATION AND DHR REVIEW WILL BE CONDUCTED. (NWV) RETAIN 5-9-2024: INCOMING RETAIN FOR ITEM# 759820 BATCH# 100322 IS NOT AVAILABLE FOR REVIEW AS PER (B)(4) WHICH STATES RETENTION SAMPLES MUST BE MAINTAINED FOR A MINIMUM OF SIX MONTHS (RECEIVED 10-4-2022). (NWV DHR 5-9-2024: RECEIVING INSPECTION FOR ITEM# 759820 BATCH# 100322 HAS BEEN ATTACHED TO THIS CASE. ALL INCOMING SUPPLIER DOCUMENTATION/CERTIFICATIONS MEET SPECIFICATIONS AS WELL AS ALL VISUAL, DIMENSIONAL, AND FUNCTIONAL TESTS AS PER 0290-IP-7.5-80-15. A TOTAL OF (B)(4) PIN TWEEZERS WAS RECEIVED ON 10-4-2022 AND AQL SAMPLING SIZE FOR VISUAL WAS (B)(4) PCS AND FUNCTIONAL EACH TEST WAS (B)(4) PCS, WITH ALL INDIVIDUAL PIN TWEEZERS MEETING ALL SPECIFICATIONS PER PROCEDURE. REVIEW OF THE INSPECTION RESULTED IN NO ISSUES AND ALL INSPECTIONS WERE PERFORMED ON THE LOT AND DEEMED ACCEPTABLE BY QUALITY. (NWV).
IN THIS EVENT IT IS REPORTED THAT PALODENT PLUS PIN TWEEZERS - THE PRONG BROKE OFF OF THE NEW PALODENT TWEEZERS. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1406321 | PALODENT PLUS PIN TWEEZERS | INSTRUMENTS, DENTAL HAND | DZN | DENTSPLY LLC | ZZ600L (DON'T THINK THIS IS RI | D0026598201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |