FDA Adverse Event Malfunction Summary report: N

PALODENT PLUS PIN TWEEZERS

MDR report key: 19091694 · Received April 11, 2024

Report

Report Number
2515379-2024-00018
Event Type
Malfunction
Date Received
April 11, 2024
Report Date
May 10, 2024
Manufacturer
DENTSPLY LLC
Product Code
DZN
UDI-DI
D0026598201
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

INVESTIGATION RETURN 5-9-2024: RETURNED PRODUCT 1 PAIR OF PALODENT PLUS PIN TWEEZERS ITEM# 759820 BATCH# 100322 WAS INSPECTED AND CONFIRMED TO BE MISSING THE ¿PIN¿ AT THE TIPS OF THE TWEEZERS WHICH AIDS IN HOLDING THE TAB MATRICES IN PLACE WITHIN THE INSTRUMENT. THE RETURNED PRODUCT IS NOT TO SPECIFICATION AND DHR REVIEW WILL BE CONDUCTED. (NWV) RETAIN 5-9-2024: INCOMING RETAIN FOR ITEM# 759820 BATCH# 100322 IS NOT AVAILABLE FOR REVIEW AS PER (B)(4) WHICH STATES RETENTION SAMPLES MUST BE MAINTAINED FOR A MINIMUM OF SIX MONTHS (RECEIVED 10-4-2022). (NWV DHR 5-9-2024: RECEIVING INSPECTION FOR ITEM# 759820 BATCH# 100322 HAS BEEN ATTACHED TO THIS CASE. ALL INCOMING SUPPLIER DOCUMENTATION/CERTIFICATIONS MEET SPECIFICATIONS AS WELL AS ALL VISUAL, DIMENSIONAL, AND FUNCTIONAL TESTS AS PER 0290-IP-7.5-80-15. A TOTAL OF (B)(4) PIN TWEEZERS WAS RECEIVED ON 10-4-2022 AND AQL SAMPLING SIZE FOR VISUAL WAS (B)(4) PCS AND FUNCTIONAL EACH TEST WAS (B)(4) PCS, WITH ALL INDIVIDUAL PIN TWEEZERS MEETING ALL SPECIFICATIONS PER PROCEDURE. REVIEW OF THE INSPECTION RESULTED IN NO ISSUES AND ALL INSPECTIONS WERE PERFORMED ON THE LOT AND DEEMED ACCEPTABLE BY QUALITY. (NWV).

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT PALODENT PLUS PIN TWEEZERS - THE PRONG BROKE OFF OF THE NEW PALODENT TWEEZERS. NO INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1406321 PALODENT PLUS PIN TWEEZERS INSTRUMENTS, DENTAL HAND DZN DENTSPLY LLC ZZ600L (DON'T THINK THIS IS RI D0026598201

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown