FDA Adverse Event Death Summary report: N

TOTALCARE

MDR report key: 1908887 · Received November 19, 2010

Report

Report Number
1824206-2010-40945
Event Type
Death
Date Received
November 19, 2010
Date of Event
October 25, 2010
Report Date
October 26, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE CPR LINKAGE WAS OUT OF ADJUSTMENT. THE TECHNICIAN ADJUSTED THE CPR LINKAGE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THAT A PT CODED IN THE BED WHEN THE CPR WAS USED. IT TOOK APPROXIMATELY 22 SECONDS TO LOWER THE HEAD SECTION BEFORE THE STAFF COULD ADMINISTER CPR. THE PT DID EXPIRE, BUT IT CANNOT BE DETERMINED IF THE HEAD SECTION LOWERING SLOWLY HAD ANY IMPACT ON THE PATIENT'S OUTCOME. THE BED WAS LOCATED IN (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1 UNK Death