FDA Adverse Event Malfunction Summary report: N

3080 SP SURGICAL TABLE

MDR report key: 19087811 · Received April 11, 2024

Report

Report Number
1043572-2024-00021
Event Type
Malfunction
Date Received
April 11, 2024
Date of Event
March 13, 2024
Report Date
April 11, 2024
Manufacturer
STERIS CORPORATION - MONTGOMERY
Product Code
FWW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE TABLE AND OBSERVED A BUILDUP OF RUST AROUND THE TABLE'S CASTERS WHICH MADE THE TABLE DIFFICULT TO MOVE. WHILE ONSITE THE TECHNICIAN REPLACED ALL FOUR OF THE SURGICAL TABLE'S CASTERS, TESTED THE SURGICAL TABLE, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. THE TECHNICIAN COUNSELED THE USER FACILITY REGARDING A PREVENTIVE MAINTENANCE PROGRAM TO ENSURE ROUTINE CHECKS FOR DAMAGED OR WORN PARTS. THE TABLE WAS MANUFACTURED IN 1997 AND IS 27 YEARS OLD. THE SURGICAL TABLE IS NOT UNDER STERIS SERVICE CONTRACT FOR MAINTENANCE. IT IS UNKNOWN WHO IS RESPONSIBLE FOR COMPLETING PREVENTIVE MAINTENANCE ACTIVITIES FOR THE SURGICAL TABLE SUBJECT OF THE EVENT. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Additional Manufacturer Narrative · 0

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE TABLE AND OBSERVED A BUILDUP OF RUST AROUND THE TABLE'S CASTERS WHICH MADE THE TABLE DIFFICULT TO MOVE. WHILE ONSITE THE TECHNICIAN REPLACED ALL FOUR OF THE SURGICAL TABLE'S CASTERS, TESTED THE SURGICAL TABLE, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. THE TABLE WAS MANUFACTURED IN 1997 AND IS 27 YEARS OLD. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem · 0

THE USER FACILITY REPORTED THAT AN EMPLOYEE STRAINED HIS BACK AND SHOULDER WHILE MOVING A 3080 SP SURGICAL TABLE. IT IS UNKNOWN IF MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
683354 3080 SP SURGICAL TABLE SURGICAL TABLE FWW STERIS CORPORATION - MONTGOMERY

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown