MOBILE INFANT WARMER
Report
- Report Number
- 9611451-2010-00720
- Event Type
- Malfunction
- Date Received
- November 25, 2010
- Report Date
- November 2, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE RETURNED WARMER HEAD HOUSING WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: THE SURFACE OF THE CASE WAS CRAZED AND THERE WERE CRACKS ON THE FRONT AND MIDDLE OF THE CASE. THE LOWER EDGE OF THE WARMER HEAD CASE WAS ALSO FOUND TO BE FLAKING OFF. DISCOLORATION WAS OBSERVED ON THE UPPER HOUSING HARNESS. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 030109. CONCLUSION: THE CAUSE OF THE CRACKING AND FLAKING IS LIKELY TO BE A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC OF THE INFANT WARMER. THE INFANT WARMER UNIT IS APPROXIMATELY 7 YEARS OLD. THE INFANT WARMER SERVICE MANUAL FOR THE AFFECTED UNITS FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES: CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTARALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE. CAUTION: THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES. OUR CLEANING INSTRUCTIONS RECOMMEND SPECIFIC PROPRIETARY CLEANING WIPES THAT CAN BE USED TO CLEAN THE FRONT PANEL FASCIAS AND OTHER HIGHLIGHTED WARMER COMPONENTS.
A HOSPITAL IN (B)(6) REPORTED THAT AN IW910 INFANT WARMER HAD A CRACKED WARMER HEAD. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MOBILE INFANT WARMER | FMT | FMT | FISHER & PAYKEL HEALTHCARE LIMITED | IW910JEU | 030109 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |