FDA Adverse Event Injury Summary report: N

GMK-REVISION 02.07.0310SCF FIXED TIBIAL INSERT SC SIZE 3/10MM

MDR report key: 19077327 · Received April 10, 2024

Report

Report Number
3005180920-2024-00192
Event Type
Injury
Date Received
April 10, 2024
Date of Event
March 11, 2024
Report Date
April 10, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030817076
PMA / PMN Number
K103170
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 18-MAR-2024. LOT 2238093: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 29-NOV-2022. EXPIRATION DATE: 2027-11-09. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. OTHER DEVICES INVOLVED: GMK-REVISION 02.07.0683L REVISION FIXED TIBIAL TRAY CEMENTED SIZE 3 L (K123721) LOT 2215011: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-OCT-2022. EXPIRATION DATE: 2027-09-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. GMK-REVISION 02.07.2403L FEMUR REVISION PS SIZE 3 L (K102437) LOT 2207826: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25-AUG-2022. EXPIRATION DATE: 2027-07-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL THE ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT ABOUT 1 YEAR AFTER MEDACTA PRIMARY SURGERY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON REMOVED ALL COMPONENTS, PERFORMED A WASHOUT, AND REIMPLANTED PERMANENT HARDWARE. THE SURGERY WAS COMPLETED SUCCESSFULLY. PATIENT HAD COMPETITOR COMPONENTS PRIOR TO MEDACTA REVISION SYSTEM IMPLANTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
413306 GMK-REVISION 02.07.0310SCF FIXED TIBIAL INSERT SC SIZE 3/10MM KNEE FIXED TIBIAL INSERT JWH MEDACTA INTERNATIONAL SA 07630030817076

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention