FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19077093 · Received April 10, 2024

Report

Report Number
9610877-2024-52664
Event Type
Malfunction
Date Received
April 10, 2024
Date of Event
March 19, 2024
Report Date
April 10, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
PMA / PMN Number
K172156
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD SIGNAL WIRE RUPTURE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD SIGNAL WIRE. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE LIGHT GUIDE CABLE COATING DAMAGE, THE LIGHT GUIDE CABLE FOR CONTROL BODY COATING DAMAGE, THE LIGHT GUIDE CABLE FOR PRONG COATING DAMAGE, THE ROOT BRACE RUBBER (INSERTION FLEXIBLE TUBE) CRACKED, THE BENDING RUBBER DIRTY, THE LG CONNECTOR LOOSE, THE INSERTION FLEXIBLE TUBE PERFORATED UNDER AND THE ROOT BRACE; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419351 PENTAX VIDEO NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL8-J10

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown