FDA Adverse Event
Injury
Summary report: N
T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY (WITH T:CONNECT MOBILE APP)
MDR report key: 19075647
·
Received April 9, 2024
Report
- Report Number
- 3013756811-2024-58862
- Event Type
- Injury
- Date Received
- April 9, 2024
- Date of Event
- March 23, 2024
- Report Date
- April 9, 2024
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00389152000091
- PMA / PMN Number
- K203234
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
THE PUMP USER GUIDE STATES, "TANDEM DIABETES CARE, INC. RECOMMENDS PERIODICALLY CHECKING THE BATTERY LEVEL INDICATOR, CHARGING THE PUMP FOR A SHORT PERIOD OF TIME EVERY DAY (10 TO 15 MINUTES), AND ALSO AVOIDING FREQUENT FULL DISCHARGES." NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE CUSTOMER ALLOWED THE PUMP BATTERY TO FULLY DEPLETE DUE TO NOT CHARGING THE BATTERY AND THE PUMP SUBSEQUENTLY SHUT OFF. CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS LOW, HOWEVER, A SPECIFIC VALUE WAS NOT PROVIDED. CUSTOMER CONSUMED CARBOHYDRATES TO ADDRESS BG LEVEL. REPORTEDLY THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2288866 | T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY (WITH T:CONNECT MOBILE APP) | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00389152000091 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Male |