FDA Adverse Event Malfunction Summary report: N

I-STAT EG7+ CARTRIDGE

MDR report key: 19073952 · Received April 9, 2024

Report

Report Number
2245578-2024-00078
Event Type
Malfunction
Date Received
April 9, 2024
Date of Event
March 17, 2024
Report Date
June 4, 2024
Manufacturer
ABBOTT POINT OF CARE
Product Code
CHL
PMA / PMN Number
K940918
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 28-MAY-2024. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AN, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR EG7+ CARTRIDGE LOT N23305, CHEM8+ CARTRIDGE LOT H23287 OR EC8+ CARTRIDGE LOT K24009.

Description of Event or Problem · 0

ON 03-APR-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG7+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT POTASSIUM RESULT ON AN 84 YEAR OLD FEMAL PATIENT WITH SEPSIS AND AGE WITH ELECTROLYTE IMBALANCE. THERE WAS NO ADDITIONA PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, DATE: (B)(6) 2024, COLLECTED: 15:27, TESTED: 15:29, RESULTS: < 2.0 MMOL/L, SAMPLE: VENOUS #1. METHOD: LAB, DATE: (B)(6) 2024, COLLECTED: 15:27, TESTED: N/A, RESULTS: 4.4 MMOL/L, SAMPLE: VENOUS #2. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Description of Event or Problem · 0

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
539219 I-STAT EG7+ CARTRIDGE EG7+ CARTRIDGE CHL ABBOTT POINT OF CARE NA N23305

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown