INFANT DUAL HEATED WITH EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2010-00713
- Event Type
- Malfunction
- Date Received
- November 23, 2010
- Date of Event
- October 24, 2010
- Report Date
- October 25, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K034026
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: ONE COMPLAINT EXPIRATORY LIMB AND ONE COMPLAINT BREATHING CIRCUIT KIT WAS RETURNED UNSEALED. BOTH WITH LOT NUMBER 100420. THE RETURNED COMPLAINT DEVICES WERE VISUALLY INSPECTED, PRESSURE TESTED AND SUBMERGED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: NO DAMAGES WERE OBSERVED ON THE COMPLAINT EXPIRATORY LIMB AND THE PRESSURE TEST FOUND THE PRESSURE DROP TO BE WITHIN SPECIFICATION FOR THIS PRODUCT. THE VISUAL INSPECTION REVEALED NO DAMAGES TO THE RETURNED BREATHING CIRCUIT KIT. THE PRESSURE TEST FOUND THE PRESSURE DROP TO BE OUTSIDE THE SPECIFICATION FOR THIS PRODUCT, AND THE WATER BATH TEST IDENTIFIED THE LEAK TO BE AROUND THE SWIVEL OF THE BREATHING CIRCUIT. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: BREATHING CIRCUITS ARE COMPOSED OF MANY PARTS, THEREFORE LEAKS CAN OCCUR AT ANY OF THE CONNECTIONS. THE TWO PARTS THAT MAKE UP THE Y-SWIVEL ARE HELD TOGETHER BY A SNAP-FIT, WHICH HAS SOME INHERENT LEAKAGE THAT IS DETECTED AND COMPENSATED FOR BY THE VENTILATOR. IT IS POSSIBLE THAT IF NOT PROPERLY LOCKED IN PLACE, THE LEAK AT THE SWIVEL JOINT CAN BE ENHANCED DURING TRANSPORT OR SETUP DESPITE HAVING PASSED THE LEAK TEST AT THE TIME OF PRODUCTION. ALL CIRCUITS ARE PRESSURE TESTED BEFORE THEY ARE ALLOWED TO LEAVE THE PRODUCTION LINE. THIS IS AN AUTOMATED PROCESS AND THE CIRCUIT WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION. (B)(4).
(B)(4). WE ARE CURRENTLY ENDEAVOURING TO OBTAIN MORE INFORMATION WITH REGARD TO THE COMPLAINT DEVICES. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED THAT ON THE (B)(6) 2010 AN RT236 INFANT DUAL HEATED EVAQUA BREATHING CIRCUIT FAILED A VENTILATOR LEAK TEST. SUBSEQUENTLY ON THE (B)(6) 2010, THE HOSPITAL REPORTED THAT ANOTHER RT236 ALSO FAILED THE VENTILATOR LEAK TEST. THE HOSPITAL FURTHER REPORTED THAT IN BOTH CASES WHEN THE CIRCUITS WERE CHANGED, THE TESTS WERE SUCCESSFUL. THIS WAS FOUND PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED THAT ON (B)(6) 2010 AN RT236 INFANT DUAL HEATED EVAQUA BREATHING CIRCUIT FAILED A VENTILATOR LEAK TEST. SUBSEQUENTLY, ON (B)(6) 2010, THE HOSPITAL REPORTED THAT ANOTHER RT236 ALSO FAILED THE VENTILATOR LEAK TEST. THE HOSPITAL FURTHER REPORTED THAT IN BOTH CASES WHEN THE CIRCUITS WERE CHANGED, THE TESTS WERE SUCCESSFUL. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT DUAL HEATED WITH EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT236 | 100420 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SERVO-I VENTILATOR| SERVO-I VENTILATOR |