FDA Adverse Event Malfunction Summary report: N

RESPIRONICS

MDR report key: 19059810 · Received April 8, 2024

Report

Report Number
2518422-2024-17966
Event Type
Malfunction
Date Received
April 8, 2024
Date of Event
March 14, 2024
Manufacturer
RESPIRONICS, INC.
Product Code
MNT
PMA / PMN Number
K102985
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H11: G1 PHONE NUMBER (B)(6).

Additional Manufacturer Narrative · 0

THE CUSTOMER INFORMED THE REMOTE SERVICE ENGINEER (RSE) OF THE ISSUE AND REQUESTED THE PART NUMBERS REQUIRED FOR A REPAIR. THE RSE PROVIDED THE CUSTOMER WITH THE PART INFORMATION FOR THE 2ND GENERATION LIQUID CRYSTAL DISPLAY (LCD), 2ND GENERATION LCD CABLE, 2ND GENERATION USER INTERFACE (UI) PRINTED CIRCUIT BOARD ASSEMBLY (PCBA), LCD TO UI PCBA CABLE, 2ND GENERATION LCD TRAY, AND SCREW PACK. MULTIPLE GOOD FAITH EFFORTS (GFE) WERE ATTEMPTED TO OBTAIN CONFIRMATION OF THE PART ORDER, PART REPLACEMENT, AND RESOLUTION. NO RESPONSE OR FURTHER INFORMATION WAS RECEIVED FROM THE CUSTOMER. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 0

PHILIPS RECEIVED A COMPLAINT ON THE V60 VENTILATOR INDICATING THAT THE SCREEN WAS VERY DIM AND ADJUSTMENTS TO THE BRIGHTNESS DID NOT HELP. THE DEVICE WAS REPORTED TO BE OUTSIDE OF USE AT THE TIME OF THE REPORTED PROBLEM. NO PATIENT OR USER HARM REPORTED. THE CUSTOMER INFORMED THE REMOTE SERVICE ENGINEER (RSE) OF THE ISSUE AND REQUESTED THE PART NUMBERS REQUIRED FOR A REPAIR. THE RSE PROVIDED THE CUSTOMER WITH THE PART INFORMATION FOR THE 2ND GENERATION LIQUID CRYSTAL DISPLAY (LCD), 2ND GENERATION LCD CABLE, 2ND GENERATION USER INTERFACE (UI) PRINTED CIRCUIT BOARD ASSEMBLY (PCBA), LCD TO UI PCBA CABLE, 2ND GENERATION LCD TRAY, AND SCREW PACK. THIS INVESTIGATION IS ONGOING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
120689 RESPIRONICS VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE MNT RESPIRONICS, INC. V60 V60PLUS VENTILATOR

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown