FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 19056999 · Received April 7, 2024

Report

Report Number
9617032-2024-00520
Event Type
Malfunction
Date Received
April 7, 2024
Date of Event
March 8, 2024
Report Date
April 17, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
FPA
UDI-DI
00382903673445
PMA / PMN Number
K220212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES. D.9:DEVICE EVAL BY MANUFACTURER? YES. D9: RETURNED TO MANUFACTURER ON: 2024-APRIL- 15TH. H.6 INVESTIGATION SUMMARY MATERIAL #: 367344. LOT/BATCH #: 3173409. BD RECEIVED 2 SAMPLES FOR INVESTIGATION. THE SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR CRACKED LUER ADAPTER WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF CRACKED LUER ADAPTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE CRACKED LUER ADAPTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.

Additional Manufacturer Narrative · 0

D2A. COMMON DEVICE NAME: BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET D2B. MEDICAL DEVICE TYPE: JKA H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, THE LUER CONE IS CRACKED, AND THERE IS A SMALL DRIP WHEN DRAWING BLOOD. THIS OCCURRED IN TEN (10) DEVICES. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, THE LUER CONE IS CRACKED, AND THERE IS A SMALL DRIP WHEN DRAWING BLOOD. THIS OCCURRED IN TEN (10) DEVICES. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1084254 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET SEE H.10 FPA BECTON, DICKINSON AND COMPANY (BD) 3173409 00382903673445

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown