BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET
Report
- Report Number
- 9617032-2024-00520
- Event Type
- Malfunction
- Date Received
- April 7, 2024
- Date of Event
- March 8, 2024
- Report Date
- April 17, 2024
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- FPA
- UDI-DI
- 00382903673445
- PMA / PMN Number
- K220212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES. D.9:DEVICE EVAL BY MANUFACTURER? YES. D9: RETURNED TO MANUFACTURER ON: 2024-APRIL- 15TH. H.6 INVESTIGATION SUMMARY MATERIAL #: 367344. LOT/BATCH #: 3173409. BD RECEIVED 2 SAMPLES FOR INVESTIGATION. THE SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR CRACKED LUER ADAPTER WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF CRACKED LUER ADAPTER WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE CRACKED LUER ADAPTER. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.
D2A. COMMON DEVICE NAME: BLOOD SPECIMEN COLLECTION DEVICE; INTRAVASCULAR ADMINISTRATION SET D2B. MEDICAL DEVICE TYPE: JKA H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, THE LUER CONE IS CRACKED, AND THERE IS A SMALL DRIP WHEN DRAWING BLOOD. THIS OCCURRED IN TEN (10) DEVICES. NO PATIENT IMPACT REPORTED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET, THE LUER CONE IS CRACKED, AND THERE IS A SMALL DRIP WHEN DRAWING BLOOD. THIS OCCURRED IN TEN (10) DEVICES. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1084254 | BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET | SEE H.10 | FPA | BECTON, DICKINSON AND COMPANY (BD) | 3173409 | 00382903673445 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |