FDA Adverse Event Malfunction Summary report: N

ARCHITECT I2000SR ANALYZER

MDR report key: 1905647 · Received November 23, 2010

Report

Report Number
1628664-2010-00433
Event Type
Malfunction
Date Received
November 23, 2010
Date of Event
October 29, 2010
Report Date
November 9, 2010
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K983212
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). TO INVESTIGATE THE ISSUE, THE COMPLAINT TEXT, THE INSTRUMENT HISTORY, AND ARCHITECT SYSTEM LABELING WERE REVIEWED. AFTER FIELD SERVICE REPLACED THE SAMPLE PROBE, THE INSTRUMENT RETURNED TO RUNNING WITHIN SPECIFICATIONS. NO ADDITIONAL OCCURRENCES OF DISCREPANT SYPHILIS TP RESULTS HAVE BEEN OBSERVED. THE REVIEW SHOWED THE PRODUCT LABELING DOES ADDRESS OPERATIONAL PRECAUTIONS, LIMITATIONS, AND ERRATIC RESULTS INCLUDING PROBABLE CAUSES AND CORRECTIVE ACTIONS. NO ADVERSE TREND WAS IDENTIFIED RELATED TO THIS ISSUE. THE SERVICE HISTORY REVIEW DID NOT IDENTIFY ANY PREVIOUS INCIDENTS OF THE ARCHITECT I2000SR, (B)(4), GENERATING DISCREPANT RESULTS OR IMPRECISE CONTROLS DURING THIS TIME FRAME. NO ADDITIONAL ERRATIC OCCURRENCES WERE OBSERVED SINCE THE COMPONENT REPLACEMENT OF THE SAMPLE PROBE. BASED ON THE AVAILABLE INFORMATION, NO PRODUCT DEFICIENCY WAS DETERMINED.

Additional Manufacturer Narrative · 1

(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATES THAT A PATIENT SPECIMEN PROCESSED USING THE ARCHITECT SYPHILIS TP ASSAY GENERATED A FALSELY (B)(6). THE SAMPLE WAS TESTED USING AN ALTERNATE METHOD (TPHA) YIELDING A (B)(6). THE SAMPLE WAS ALSO REPEATED ON THE ARCHITECT YIELDING A (B)(6). NO ADVERSE OUTCOMES WERE REPORTED RELATED TO THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT I2000SR ANALYZER AUTOMATED IMMUNOASSAY ANALYZER JJE ABBOTT MANUFACTURING, INC.

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT SYPHILIS TP LN 8D06-20 LOT # UNK| ARCHITECT SYPHILIS TP LN 8D06-20 LOT # UNK