FDA Adverse Event Malfunction Summary report: N

ANGLED REAMER, DRIVE SHAFT

MDR report key: 19056056 · Received April 5, 2024

Report

Report Number
1220246-2024-01853
Event Type
Malfunction
Date Received
April 5, 2024
Date of Event
March 2, 2023
Report Date
April 18, 2024
Manufacturer
ARTHREX, INC.
Product Code
LXH
UDI-DI
00888867324954
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

THIS 3500A RECORD IS SUBMITTED TO COMPLY WITH AN FDA 483 INSPECTIONAL OBSERVATION ISSUED TO ARTHREX INC ON MAY 5, 2023. ARTHREX HAS REASSESSED THE REPORTABILITY DECISIONS MADE ON HISTORICAL COMPLAINT RECORDS USING REVISED CRITERION. THIS 3500A DOCUMENT IS A RESULT OF THE REASSESSMENT. COMPLAINT IS CONFIRMED. VISUAL EVALUATION FOUND HEAVY CIRCULAR SCRATCHES AROUND THE OUTSIDE DIAMETER AT THE DISTAL END OF THE SHAFT AND CHIPPED. THE PROBABLE CAUSE IS WHEN THE MATTING METAL PARTS WERE ACTIVATED. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO MISUSE/MISHANDLING DUE TO DAMAGE TO THE DEVICE.

Additional Manufacturer Narrative · 0

THIS 3500A RECORD IS SUBMITTED TO COMPLY WITH AN FDA 483 INSPECTIONAL OBSERVATION ISSUED TO ARTHREX INC ON MAY 5, 2023. ARTHREX HAS REASSESSED THE REPORTABILITY DECISIONS MADE ON HISTORICAL COMPLAINT RECORDS USING REVISED CRITERION. THIS 3500A DOCUMENT IS A RESULT OF THE REASSESSMENT. COMPLAINT IS CONFIRMED. VISUAL EVALUATION FOUND HEAVY CIRCULAR SCRATCHES AROUND THE OUTSIDE DIAMETER AT THE DISTAL END OF THE SHAFT AND CHIPPED. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO WEAR AND TEAR. H6, H10.

Description of Event or Problem · 0

ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-9597-10 ANGLED REAMER SLEEVE, AND AN AR-9676 DRIVE SHAFT ARE FUSED TOGETHER. THIS OCCURRED DURING A CASE, WITH NO PATIENT EFFECT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1405618 ANGLED REAMER, DRIVE SHAFT MANUAL INSTR, GENERAL SURGICAL LXH ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT 022213 00888867324954

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown