FDA Adverse Event Injury Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 1905582 · Received November 19, 2010

Report

Report Number
2028159-2010-02228
Event Type
Injury
Date Received
November 19, 2010
Date of Event
October 11, 2010
Report Date
October 27, 2010
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. CORNEAL BURN IS AN ISSUE THAT IS OCCASIONALLY REPORTED WITH CATARACT SURGERY. ACCORDING TO THE (B)(6) HEALTH DEVICES, HAZARD UPDATE: SCLERAL AND CORNEAL BURNS DURING PHACOEMULSIFICATION, NOVEMBER 1996, VOL. 25, NO. 11: 426-431, MOST CORNEAL BURNS CAN BE TRACED TO ISSUES RELATED TO SURGICAL TECHNIQUE AND NOT TO MALFUNCTIONING EQUIPMENT. A LETTER AND COPY OF THIS INDUSTRY ARTICLE WAS PROVIDED TO THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE SURGEON REPORTED A CORNEAL BURN OCCURRED. DURING CATARACT SURGERY OF THE LEFT EYE, EVERYTHING WENT SMOOTHLY DURING THE FIRST STEPS OF THE PROCEDURE. WHEN THE SURGEON WAS SCULPTING THE EPICORTEX AND THE BEGINNING OF THE FIRST GROOVE, THE ASPIRATION SUDDENLY STOPPED. A WHITISH VISCOUS PLUG WAS OBSERVED. THE SURGEON THOUGHT HE COULD ASPIRATE IT, SO HE PRESSED DOWN ON THE FOOTSWITCH. ONLY THE PHACO ULTRASOUND WAS AVAILABLE. THERE WAS NO IRRIGATION/ASPIRATION. THE CORNEAL BURN LEAD TO SEIDEL AND SECONDARY HERNIA OF THE IRIS. THE WOUND WAS SUTURED WITH ONE STITCH. ANOTHER SUTURE WAS PLACED 5 DAYS AFTER SURGERY. THE INCISION IS STILL NOT TIGHT. THE WOUND IS HEALING SLOWLY. THE PT REQUIRED HOSPITALIZATION. ADDITIONAL INFORMATION RECEIVED FROM THE SURGEON STATED THE NUCLEUS WAS NOT PARTICULARLY HARD. THE INCISION SIZE 2.2MM. THE SURGEON WAS USING 100% US PHACO POWER TO REMOVE THE VISCOUS PLUG. THE IRRIGATION SOLUTION WAS PREVIOUSLY ICED. THE SURGEON DID NOT CREATE A WORKING SPACE BY REMOVING SOME VISCOELASTIC BEFORE BEGINNING PHACO. THE PT HAD AN AMNIOTIC MEMBRANE GRAFT. THE SURGEON SAW THE PT ON (B)(6) 2010. THE PT'S PROGNOSIS IS STILL UNK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1 NI Hospitalization| R