FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM

MDR report key: 1905553 · Received November 23, 2010

Report

Report Number
3005075853-2010-06650
Event Type
Malfunction
Date Received
November 23, 2010
Date of Event
September 14, 2010
Report Date
September 16, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT DEVICE (A) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIP AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE ANALYSIS RESULTS OF DEVICE (B) FOUND THAT IT WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING THE JAWS REMAIN IN THE CLOSED POSITION, 1 PEAR SHAPED CLIPS WAS RELEASED DUE TO AND ADVANCER BYPASS, THE REMAINING 4 CLIPS WERE FIRED CONFORMING. THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS NOT VISIBLE. AS NOT ENOUGH INFORMATION ABOUT THE INCIDENT REPORTED WAS AVAILABLE, THE EVENT COULD NOT BE CONFIRMED. BATCH # G9KK3Z, MFG DATE: 07/16/2010; EXP DATE: 06/16/2015. 3500A CODES: (B)(4): ADVANCER BYPASS - MALFORMED CLIP - JAWS CLOSED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY PROCEDURE, JAMMING OCCURRED AND THE DEVICE COULD NOT BE USED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1