ECHELON*FLEX45
Report
- Report Number
- 3005075853-2010-06647
- Event Type
- Malfunction
- Date Received
- November 23, 2010
- Date of Event
- October 11, 2010
- Report Date
- October 29, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051002
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SC45A DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH AN ECR45G CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED 1/10. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. THE RETURNED CARTRIDGE RELOAD WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION BY RESETTING AND RELOADING IT INTO THE DEVICE. THE DEVICE ACHIEVED ITS COMPLETE STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED AS INTENDED. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A DEVICE WITH THE INDICATOR IN THE LOCKED POSITION A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. ONCE THE DEVICE IS OPENED THE DEVICE CAN BE DE-ARTICULATED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. (B)(4)
IT WAS REPORTED THAT THE DEVICE WAS USED DURING AN THORASCOPIC PROCEDURE. THE DEVICE BECAME STUCK ON THE BRONCHUS. THEY TRIED REVERSING THE KNIFE AND STILL COULD NOT GET IT TO RELEASE. ANOTHER DEVICE WAS USED TO FORCE THE DEVICE OFF OF THE BRONCHUS. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX45 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | G4T99D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ECR45G |