FDA Adverse Event Death Summary report: N

CONSULTA

MDR report key: 1905451 · Received November 23, 2010

Report

Report Number
2647346-2010-00743
Event Type
Death
Date Received
November 23, 2010
Date of Event
August 8, 2009
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) BATTERY DEPLETION-NORMAL. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC DEPRESSION. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC DEPRESSION. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND, OUTER INSULATION COSMETIC DEPRESSION. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED BY PATIENTS LIFE PARTNER THAT SHE QUESTIONED IF THE DEVICE PROVIDED THERAPY AT THE TIME OF THE PATIENT DEATH AND THAT THE "LEAD WIRE FRIED." THE PATIENT HAD RECENTLY COMPLETED TESTING FOR HEART TRANSPLANT AND "ALL GOOD." THE PATIENT HAD COMPLAINED OF "BUZZING IN MY CHEST" BUT NOT PAIN THE NIGHT BEFORE HE DIED AND "CAN'T BREATH" IN THE EARLY AM. 911 WAS CALLED AND THE MEDICS ARRIVED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED. FOLLOW UP WITH CLINIC REVEALED THERE WAS NO "FRIED LEAD." DEVICE INTERROGATION AFTER DEATH SHOWED DEVICE FUNCTIONING FINE, DELIVERED SHOCKS, BUT PATIENT'S HEART DID NOT RESPOND. CLINIC HAS "NO CONCERNS ABOUT THE PATIENT'S DEVICE SYSTEM PERFORMANCE PRIOR TO OR AT THE TIME OF DEATH." PATIENT HAD NOT BEEN PACEMAKER DEPENDENT. LAST DEVICE CHECK WAS ONE MONTH PRIOR TO DEATH AND THERE WERE NO RHYTHM ISSUES AND DEVICE WAS FUNCTIONING FINE WITHIN NORMAL PARAMETERS.

Description of Event or Problem · 1

IT WAS REPORTED BY PATIENTS LIFE PARTNER THAT SHE QUESTIONED IF THE DEVICE PROVIDED THERAPY AT THE TIME OF THE PATIENTS DEATH AND THAT THE "LEAD WIRE FRIED." THE PATIENT HAD RECENTLY COMPLETED TESTING FOR HEART TRANSPLANT AND "ALL GOOD." THE PATIENT HAD COMPLAINED OF "BUZZING IN MY CHEST" (NOT PAIN) THE NIGHT BEFORE HE DIED AND "CAN'T BREATH" IN THE EARLY AM. 911 WAS CALLED AND THE MEDICS ARRIVED. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 65 YR Death| H| L| R