FDA Adverse Event Malfunction Summary report: N

PHILIPS HOLTER SOFTWARE

MDR report key: 19053774 · Received April 5, 2024

Report

Report Number
1218950-2024-00242
Event Type
Malfunction
Date Received
April 5, 2024
Date of Event
March 12, 2024
Report Date
May 24, 2024
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
MLO
PMA / PMN Number
K010949
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT WAS ESCALATED FOR TECHNICAL INVESTIGATION. THE ZHR FILE FORMAT WAS DESIGNED IN AN EARLY RELEASE OF HOLTER SOFTWARE, AND THERE IS NO RESERVED SPACE FOR NEW DATA, SUCH AS THE DICOM MODALITY WORKLIST (MWL) DATA. WHEN ¿DICOM ENCAPSULATED PDF¿ WAS ADDED IN HOLTER 3.0.4 RELEASE, THE ZHR FILE FORMAT WAS NOT EXPANDED TO STORE DICOM DATA. THEREFORE, TO AVOID LOSING DICOM MWL DATA, THE CORRECT DICOM WORKFLOW WAS DETERMINED BY CHOOSING A DICOM MWL IN ¿NEW PATIENT¿ DIALOG AFTER ECG DATA IS UPLOADED FROM DTXT RECORDER AND IS ANALYZED BY CALG ALGORITHM. IT SHOULD BE EXPORTED TO DICOM SERVER DIRECTLY. IN THE ¿DICOM ENCAPSULATED PDF¿, THE PATIENT DATA ON PDF REPORT IS CORRECT. IN OTHER WORDS, WHEN THE USER OPENS, VIEWS, AND PRINTS THE DICOM ENCAPSULATED PDF REPORT, ALL THE DATA IS CORRECT. ONLY THE DATA STORED IN DICOM TAGS IS WRONG. THE RECOMMENDED WORKFLOW IS TO EXPORT DICOM PDF REPORT FIRST, THEN SAVE TO ZHR FILE. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED, THE CAUSE OF THE REPORTED PROBLEM WAS THE DESIGN ISSUE OF SOFTWARE. THE REPORTED PROBLEM WAS CONFIRMED. E1 REPORTING ADDRESS POSTAL CODE: (B)(6). THE DICOM INFO MUST BE DELETED IN MEMORY WHEN THE USER CLOSES THE DOCUMENT. PHILIPS WILL CHANGE THE ZHR FILE FORMAT TO STORE DICOM MWL DATA. AFTERWARDS, THE MISMATCH WILL NOT HAPPEN ANYMORE. THE ISSUE WILL BE FULLY RESOLVED IN A FUTURE SOFTWARE RELEASE.

Additional Manufacturer Narrative · 0

PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THERE IS A PATIENT MISMATCH IN CLINICAL HOLTER (ECG) REPORT DATA BETWEEN TWO PATIENTS IN THE PHILIPS CARDIO INTELLISPACE CARDIOVASCULAR APPLICATION. THE RESULT IS THAT THE WRONG PATIENT DATA IS MATCHED WITH THE WRONG PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1448633 PHILIPS HOLTER SOFTWARE PHILIPS HOLTER SOFTWARE MLO PHILIPS MEDICAL SYSTEMS 860292

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown