NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2010-00708
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Date of Event
- October 28, 2010
- Report Date
- October 28, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). METHOD: THE COMPLAINT RD900AWU NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO FPH SERVICE CENTER IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR DAMAGE AND PERFORMANCE TESTED FOR THE REPORTED FAULT. RESULTS: THE REPORTED FAULT COULD NOT BE REPLICATED AS THE VALVE SYSTEM AND THE MANOMETER OPERATED PROPERLY DURING PERFORMANCE TESTING. HOWEVER, VISUAL INSPECTION SHOWED THAT THE GAS INLET CONNECTOR WAS BROKEN AND THE NEOPUFF CASING WAS CRACKED. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. IT CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF". FURTHERMORE, THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "THE INTEGRITY OF THE SYSTEM AND MANOMETER SHOULD BE CHECKED PRIOR TO FIRST USE, ANNUAL AND AFTER SERVICING", "WARNING DROPPING OF THE F&P NEOPUFF INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS [AS OUTLINED IN THE MANUAL] BEFORE CONNECTION TO A PATIENT".
(B)(4). THE RD900AWU NEOPUFF INFANT RESUSCITATOR IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE AND HAVE COMPLETED OUR INVESTIGATION.
A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RD900AWU NEOPUFF INFANT RESUSCITATOR DID NOT PASS THE VALVE SYSTEM TEST. THE REQUIRED PRESSURE WAS NOT ACHIEVED WHEN THE GAS SUPPLY WAS SET TO 5 LPM, AS SPECIFIED IN THE PERFORMANCE CHECK PROCEDURE. THIS WAS NOTICED AFTER USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RD900AWU NEOPUFF INFANT RESUSCITATOR DID NOT PASS THE VALVE SYSTEM TEST. THE REQUIRED PRESSURE WAS NOT ACHIEVED WHEN THE GAS SUPPLY WAS SET TO 5 LPM, AS SPECIFIED IN THE PERFORMANCE CHECK PROCEDURE. THIS WAS NOTICED AFTER USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LIMITED | RD900AWU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |