FDA Adverse Event Malfunction Summary report: N

POWER PROCESSOR WITH LINE CONTOLLER

MDR report key: 1905215 · Received November 22, 2010

Report

Report Number
2050012-2010-01356
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
October 27, 2010
Report Date
October 27, 2010
Manufacturer
BECKMAN COULER INC.
Product Code
JQP
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

A BCI FIELD SERVICE ENGINEER (FSE) DISCOVERED THE END TRACK DID NOT TURN ON WHEN LINE WAS ON. FSE NOTICED THE GREEN BELTS WERE OFF FROM INLET TO FIRST HANDING TRACK PAST THE SMART H LINE. FSE REPLACED A BURNED CONNECTOR. FSE POWERED UP THE LINE AND ALL TRACKS BELTS WERE IN WORKING ORDER.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO TRACKS NOT MOVING ON THE AUTOMATION LINE. NO OPERATOR EXPOSURE OF INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWER PROCESSOR WITH LINE CONTOLLER CLINICAL AUTOMATION JQP BECKMAN COULER INC. N/A N/A

Patients

Seq Age Sex Outcome Treatment
1