FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 1905208 · Received November 22, 2010

Report

Report Number
2050012-2010-01354
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
October 25, 2010
Report Date
October 25, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER HAD REPLACED THE SYRINGES ABOUT 4-6 MONTHS AGO. THE CUSTOMER CONTACTED SERVICE AND THE SERVICE INSTRUCTED THE CUSTOMER TO REPLACE THE SYRINGE BARREL AND PLUNGER. THE CUSTOMER REPLACED THE PARTS AND RAN QC. NO ADDITIONAL LEAK HAS BEEN REPORTED SINCE. QA REQUESTED THE SYRINGES TO BE RETURNED BY THE CUSTOMER.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO THE UCTA SYRINGE BARREL AND PLUNGER LEAK ON THE UNICEL DXC 600I SYNCHRON CHEMISTRY ANALYZER. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment
1