FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 1905192 · Received November 22, 2010

Report

Report Number
2955842-2010-00511
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
November 3, 2010
Report Date
November 8, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING FOUND THE INSTRUMENT GRIPS EXHIBIT AN OFFSET OF APPROXIMATELY .015 INCHES AT THE TIPS WHEN CLOSED. THE YAW PULLEY WAS FOUND BROKEN AND CONDUCTOR WIRE IS BROKEN AND DETACHED FROM THE CONNECTION AT THE GRIP. THE YAW PULLEY IS MISSING A .160 INCH BY .060 INCH PIECE OPPOSITE THE CONDUCTOR BREAK ALLOWING THE GRIP TO MOVE WITHIN THE PULLEY AND HAVE A LARGER RANGE OF MOTION IN YAW. PER THE CASE NOTES, THE INSTRUMENT FRAGMENT WAS RETRIEVED FROM THE PATIENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE A PIECE OF THE PK DISSECTING FORCEPS INSTRUMENT FELL INTO THE PATIENT. THE PIECE WAS RETRIEVED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M1010323 290

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM, INSTRUMENTS, ACCESSORIES, & ES