FDA Adverse Event Malfunction Summary report: N

TRUWAVE DISPOSABLE PRESSURE TRANSDUCER

MDR report key: 1905139 · Received November 22, 2010

Report

Report Number
2015691-2010-14413
Event Type
Malfunction
Date Received
November 22, 2010
Date of Event
August 30, 2010
Report Date
September 2, 2010
Manufacturer
EDWARDS LIFESCIENCES PR
Product Code
DXO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: PRIMING SOLUTION WAS VISIBLE BUT NO BLOOD WAS FOUND INSIDE THE RETURNED DEVICES. LEAKAGES WERE CONFIRMED AROUND WELD AREA OF FLUSH DEVICES DURING LEAK TEST AT PRESSURE 300MMHG. THE CAPS WERE WELDED TO FLUSH DEVICE HOUSINGS. ENERGY LINES AROUND WELD AREA OF FLUSH DEVICE WERE MELTED. ONE CRACK, APPROXIMATELY .2" IN LENGTH, WAS ALSO FOUND AT THE WELD AREA OF BOTH FLUSH DEVICE HOUSINGS, HOWEVER LEAKAGES DID NOT OCCUR THROUGH THE CRACKS. THE REPORTED DEFECT WAS CONFIRMED TO BE A MANUFACTURING DEFECT. PRESSURE TUBING WAS FOUND COMPLETELY DETACHED FROM THE UV BOND JOINT OF THE VAMP ADULT RESERVOIR. HALF OF BROKEN TUBING WAS NOT RETURNED WITH THE SYSTEM AND THE OTHER HALF WAS STILL INSIDE RESERVOIR TUBE HOUSING. ROUGH SURFACE WAS OBSERVED AT BROKEN TUBING END. IT APPEARS THAT THE BOND JOINT WAS NOT SUFFICIENTLY SECURED DUE TO INADEQUATE UV ADHESIVE. AN INVESTIGATION WAS OPENED FOR THIS ISSUE AND MULTIPLE CORRECTIVE ACTIONS WERE IMPLEMENTED. THESE INCLUDE: IN-PROCESS CHANGES, ADJUSTMENTS OF FIXTURES, REPLACEMENT OF TYPE OF BONDING, UPDATED PREVENTATIVE MAINTENANCE, AND NEW PROCESS SPECIFICATIONS/ALERT LIMITS FOR PULL STRENGTH VALUES. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED AND DOCUMENTED THAT THE DEVICE MET ALL SPECIFICATIONS UPON DISTRIBUTION.

Description of Event or Problem · 1

REPORTEDLY, THE TRANSDUCER LEAKS. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRUWAVE DISPOSABLE PRESSURE TRANSDUCER DISPOSABLE PRESSURE TRANSDUCER DXO EDWARDS LIFESCIENCES PR CDPUNKNOWN 58780731

Patients

Seq Age Sex Outcome Treatment
1