FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2010-01588
- Event Type
- Malfunction
- Date Received
- November 22, 2010
- Date of Event
- November 2, 2010
- Report Date
- November 22, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
CUSTOMER'S METER (B)(4) WAS RETURNED AND TESTED WITH TEST STRIPS LOT # 0966918 . THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION. ADDITIONALLY, THE REPORTED READINGS OF 3.8 MMOL/L (68 MG/DL), 11.8 MMOL/L (213 MG/DL) AND 14.5 MMOL/L (261 MG/DL) WERE FOUND IN THE DEVICE'S INTERNAL MEMORY LOG WITHIN 10 MINUTES.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 3.8 MMOL/L, 11.8 MMOL/L, AND 14.5 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 0966918 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |